Preloaded spin columns with buffer-flush activation for 10-minute depletion of HSA from plasma. ProtiPrep™ Pop1 is a fully integrated, ready-to-use microspin column for the rapid depletion of human serum albumin (HSA), the most abundant protein in plasma.
Plasma is one of the most information-rich yet analytically complex biological fluids due to its extraordinary dynamic range of protein concentration.
Highly abundant proteins like human serum albumin (HSA) dominate the proteome, accounting for nearly 50 to 60 % of the total protein mass, while clinically relevant low-abundance proteins may be present at concentrations up to 12 orders of magnitude lower.
To overcome this challenge, ProtiPrep Pop1 depletes >97 % of HSA from 10 µl of plasma within 10 minutes providing minimal nonspecific interaction with other proteins.
ProtiPrep simplifies sample preparation to the maximum. Apply 10 µl of plasma (~800 µg), screw the cap shut, incubate, centrifuge and collect the sample. The prepared sample (~150 µl) is stored in PBS buffer (pH 7.4, 0.02 % sodium azide) and ready for analysis or subsequent processing. However, the amount of HSA in plasma can vary considerably. For best results, optimize the ratio of sample to slurry volume for each specific application.
Depletion of Human Serum Albumin (HSA)
Product: AMI001HSA 6/12/24 Columns
Storage Temperature: 2 – 8°C
Do not freeze.
Step 1: Equlibrate the ProtiPrep column to room temperature. Open the screw cap (blue) and apply 10 µl of plasma to the cavity.
Step 2: Screw the buffer compartment cap (yellow) onto the tube and tighten until it clicks.
Step 3: Tap the column firmly against a solid surface to fully release the buffer onto the resin.
Step 4: Shake the tube vertically by hand for at least 10 seconds and incubate at room temperature for 5 minutes.
Step 5: Centrifuge for 2 minutes at 1000 x g.
Step 6: Unscrew. The column is getting removed together with the lid. The HSA depleted samples remain in the tube. The sample may be used immediately for further processing or stored at −20 °C for later use.
| Observation | Possible cause | Recommended action |
|---|---|---|
| HSA was not completely removed. | Sample exceeded binding capacity. | Reduce the amount of sample processed. |
| Incomplete binding. | Increase the incubation time. | |
| Sample was not mixed during incubation. | Mix the sample with resin gently to allow buffer flush over properly. |
Note: Sample processing will depend on the type of downstream analysis and may require buffer exchange, lipid and other metabolite removal and/or concentration for 2D gel electrophoresis or mass spectrometry analysis after depletion.
ProtiPrep are pre-packed with the optimum amounts of functionalized agarose beads and PBS buffer (pH 7.4, 0.02 % sodium azide), to ensure efficient and reproducible sample preparation without the need for special laboratory equipment (Microcentrifuge capable of operating 1000 x g).
Plasma is one of the most information-rich yet analytically complex biological fluids due to its extraordinary dynamic range of protein concentration.
Highly abundant proteins like human serum albumin (HSA) dominate the proteome, accounting for nearly 50 to 60 % of the total protein mass, while clinically relevant low-abundance proteins may be present at concentrations up to 12 orders of magnitude lower.
To overcome this challenge, ProtiPrep Pop1 depletes >97 % of HSA from 10 µl of plasma within 10 minutes providing minimal nonspecific interaction with other proteins.
Note: No other additions or solvent exchange are required before protein depletion.
Note: Single use only. Do not reuse the resin
The depletion resin is designed for use with human plasma samples and has not been tested on any other species.